Keywordslead screenmobile radiation shieldlead apronthyroid shieldpersonnel radiation protection
Where fixed shielding stops being enough
- A shielded room protects what is outside it.
- Inside, staff work close to the equipment and close to the patient, often for long procedures.
- A mobile lead screen is placed between the person and the source at exactly that working position — during a mobile X-ray at the bedside, or in an interventional suite where the operator stands near the beam for the whole case.
Screen values differ from wall values
A screen works at a greater distance and for shorter periods than a room boundary, so its lead equivalent is normally lower than the wall lining — typically 0.5–2 mmPb.
That is not a compromise; it is what the working position requires, and it keeps the screen light enough to move. The value is set from the protection needed at that point, not from the wall's figure.
Promptmobile lead screen on castors positioned beside a patient bed during a mobile X-ray procedure, staff behind the screen, clinical documentary photography
ShotA mobile lead screen placed at the working position during a procedure.
Product02.5 Lead Screens
Promptinterventional radiology suite with a lead glass viewing panel on a mobile screen, operator working behind it, medical technology interior
ShotA screen with a lead glass viewing panel at the working position.
Product02.5 Lead Screens
Personal protection follows the same logic
Aprons, thyroid shields and gloves are the last layer. Their lead equivalent follows from the time, distance and position of the wearer, and is normally lower again — typically 0.25–0.5 mmPb.
Weight is the other half of the specification: an apron is carried by one person through a whole shift, so the value and the fit are decided together, and sizing is part of the specification rather than an afterthought.
