Lead-Lined Radiation Protection Doors

Lead-lined radiation protection doors are steel doors with an integrated lead core for CT, DR and X-ray rooms, made to the shielding value calculated for the room rather than to a fixed catalogue leaf.

  • Lead Equivalent1–3 mmPb, matched to the surrounding wall shielding
  • Leaf ConfigurationSingle leaf or double leaf
  • Clear Passage Width900–1,500 mm (single leaf); 1,400–2,400 mm (double leaf)
  • OperationManual swing, or automatic sliding / swing
  • Vision PanelLead glass, optional
Lead-Lined Radiation Protection Doors — application overview

Overview

A lead-lined radiation protection door is a steel door built around a lead core, fitted at the opening of a room that has to hold radiation inside it. The door is not chosen from a catalogue thickness — it is made to the lead equivalent value calculated for the room, so the shielding is continuous from the wall, across the frame and through the leaf.

These doors are used at CT scanning rooms, DR and X-ray rooms, interventional and angiography rooms, radiotherapy rooms, nuclear medicine rooms and veterinary imaging rooms — anywhere the room boundary is part of the shielding design rather than just a partition.

The lead core sits inside the leaf and is matched to the wall value, and the frame detail carries the shielding around the opening so there is no unshielded line at the junction between wall and door. Where staff need to see into the room, a lead glass vision panel is fitted with the same lead equivalent as the leaf, so the window does not become the weak point in the room boundary.

Selection starts from the room's shielding requirement and the door schedule: the lead equivalent, the opening size and the direction of travel. Manual or automated operation follows from the workflow — an angiography room with a heavy workload is worked differently from an X-ray room used occasionally.

Specifications

Product TypeLead-lined door for radiation-shielded rooms
Typical UseCT, DR, X-ray, interventional, radiotherapy and nuclear medicine rooms
Lead Equivalent1–3 mmPb, matched to the surrounding wall shielding
Leaf ConfigurationSingle leaf or double leaf
Clear Passage Width900–1,500 mm single leaf; 1,400–2,400 mm double leaf, customizable
Door Height2,100–2,400 mm, customizable
Leaf Thickness40–60 mm, set by the lead core and hardware detail, customizable
OperationManual swing, or automatic sliding / swing
Vision PanelLead glass, optional, at the same lead equivalent as the leaf
FrameSteel frame with shielding continued around the opening
FinishPowder-coated steel
HardwareHandle, closer, lock and stop options
PackingPlywood case with the leaf braced for transport
NotesValues shown are typical of this door type. Final specification is confirmed against the room's shielding requirement.

Features

  • Shielding matched to the room — The lead core is made to the value calculated for the room, not taken from a fixed leaf thickness.
  • Continuous at the opening — The frame detail carries the shielding across the doorway, so the junction between wall and door is not left unshielded.
  • Lead glass panel option — A vision panel at the same lead equivalent as the leaf, so the observation window does not weaken the boundary.
  • Manual or automatic — Operation chosen from the department's workload rather than from a default.
  • Leaf sized to the opening — Single or double leaf, with the weight and hardware matched so the door remains workable day to day.
  • Crated for the lead core — The leaf is braced in a plywood case, since a lead core adds weight and the edges are what suffer in transit.

Available Customization

Lead equivalent1–3 mmPb, matched to the wall shielding calculated for the room
Leaf configurationSingle or double leaf, to suit the opening and the equipment passing through
Clear passage widthSingle leaf 900–1,500 mm; double leaf 1,400–2,400 mm, made to the project's structural opening
Door height2,100–2,400 mm, made to the opening height
OperationManual swing, automatic sliding, or automatic swing
Vision panelLead glass panel at the same lead equivalent as the leaf, sized to the sightline required
FrameSteel frame with the shielding continued around the opening
HardwareHandle, closer, lock and stop options to the access requirement
FinishPowder-coated steel in the project's colour
PackingPlywood case with the leaf braced; frame labelled by opening

Applications

Lead-lined radiation protection doors are specified at the entrances to rooms where the boundary forms part of the shielding design: CT scanning rooms, DR and digital radiography rooms, X-ray examination rooms, interventional and angiography rooms, radiotherapy rooms, nuclear medicine rooms and veterinary imaging rooms. The door is selected from the room's shielding requirement and the door schedule, and its lead equivalent is normally kept the same as the surrounding wall so that the opening is not the weakest point in the room.

Lead-Lined Radiation Protection Doors — CT Room

CT Room

Lead-Lined Radiation Protection Doors — DR Room

DR Room

Lead-Lined Radiation Protection Doors — Dental X-Ray Room

Dental X-Ray Room

Lead-Lined Radiation Protection Doors — Radiotherapy Room

Radiotherapy Room

Lead-Lined Radiation Protection Doors — Industrial Inspection Room

Industrial Inspection Room

Lead-Lined Radiation Protection Doors — X-Ray Examination Room

X-Ray Examination Room

Manufacturing & Quality Control

Every unit passes the same five stages below. Checks are recorded per unit rather than sampled, and a unit does not move to the next stage until the previous one is signed off.

STEP 01

Incoming Material

Lead sheet thickness, barium sulphate density and glass lead equivalent are verified against the supplier certificate before the material enters the workshop.

STEP 02

Lead Equivalent Verification

Shielding layers are measured for thickness and uniformity at the points that matter most — corners, edges and any joint. This is the check that decides whether the product is fit to ship.

STEP 03

Joint & Overlap Inspection

Overlaps and joints are inspected for width and continuity. A gap in a shield is not a cosmetic defect, so every joint is measured rather than sampled.

STEP 04

Assembly Check

Frame, glass and shielding are checked for fit and alignment. Doors and screens are operated through their full range to confirm the shielding stays continuous in the closed position.

STEP 05

Finish & Dispatch

Surfaces and edges are inspected for damage and finish quality, then the unit is protected, labelled with its shielding data and packed against the shipping list.

Packaging & Delivery

The leaf is braced inside a plywood case so it stays true in transit, with the frame and hardware packed separately and labelled.

Plywood case
Plywood case
Protective packing
Protective packing
Product labelling
Product labelling
Export packing
Export packing
Loading
Loading
Customized packing
Customized packing

FAQ

How do we know which lead equivalent to order?

It comes from the shielding calculation for the room, which follows the equipment, the workload and the room's layout. The door is then made to that value rather than chosen from a list.

Can the door's lead equivalent be different from the wall?

It can, but it is normally kept the same. Matching the wall avoids the entrance becoming the point where the room is easiest to penetrate.

Is the shielding continuous around the frame?

Yes. The frame detail carries the lead around the opening so there is no unshielded line where the wall meets the door.

What about the vision panel?

A lead glass panel is fitted at the same lead equivalent as the leaf. Where the sightline needs a large panel, that influences the leaf design rather than being decided afterwards.

Manual or automatic — which is better?

It follows from the department's workload. Rooms with steady traffic and staff carrying equipment are usually automated; rooms used occasionally are normally manual.

How heavy does the door get?

A lead core adds weight, so the leaf weight, hinges and closer are set together. It is one reason the door is made to the opening rather than selected from a standard size.

Can it be used on a room with a mobile imaging unit?

Yes. The shielding value is calculated for the actual use of the room, so a room with mobile equipment is shielded for that workload.

What do you need from us to quote?

The room's shielding requirement, the door schedule and opening sizes, the wall build-up, and whether the door is to be manual or automatic.

Related Articles

Background reading on Lead-Lined Radiation Protection Doors and how it is specified.

Radiation Protection23 June 2026274 views

Radiation Protection Doors: Lead Core, Frame Detail and Vision Panels

A lead-lined door has to match the wall's shielding value and carry it across the opening. This article covers the lead core, the frame detail and when to add a lead glass vision panel.

Read Article
General28 May 2026294 views

How to Compare Quotations for a Healthcare Fit-Out Package

Two quotations for the same doors and shielding are rarely priced on the same scope. This article sets out what to align before comparing figures, and which differences change the cost on site.

Read Article
Radiation Protection20 April 2026189 views

X-Ray Lead Glass: Thickness, Viewing Distance and Frame Detail

An observation window carries the same lead equivalent as the wall it sits in, and its thickness depends on how far the operator stands back. This article covers what to plan before ordering.

Read Article

View All Articles

Send Us Your Project Requirements

Tell us the room type, opening size, lead equivalent requirement or door configuration. We will come back with a practical supply proposal.

Email