Working load is confirmed against the heaviest case

The working load should be set from the heaviest procedure the room will take, not from an average patient.

Tables typically run 200–350 kg. This is also the figure that decides the column and base: a table asked to hold a heavy patient at full tilt is loaded differently from one used flat, and the stability of the base is what the surgeon depends on for the whole case.

Positioning range follows the procedures

Height, Trendelenburg, reverse Trendelenburg, lateral tilt and the back section are the ranges that matter. A general theatre needs enough range to reconfigure between specialties; a dedicated room needs the range for the one type of work it does.

Typically height runs 700–1,000 mm, Trendelenburg to about 25° and lateral tilt to about 20°, with the back section adjustable from 0–80°.

Image 1 — placeholder

Promptoperating table at full lateral tilt with a patient positioning set, surgical team around it, operating theatre interior photography

ShotA table at full tilt, where base stability matters most.

Product03.3 Operating Tables

Image 2 — placeholder

Promptsurgeon adjusting an operating table using a handset control during setup, positioning controls in hand, operating theatre documentary photography

ShotTable positioning being set from the handset.

Product03.3 Operating Tables

Accessories fit to rails, so plan them with the table

  • Arm supports, leg supports, head rests and straps fit to rails on the table top, which means the accessory range is part of the specification rather than bought separately afterwards.
  • A theatre that buys the table first and the accessories later usually ends up with rails that do not take what the department needs, or a changeover that takes longer than the list allows.